Electrochemical Detector


The Electrochemical Detector (ECD) detects electroactive compounds in HPLC and IC via redox reactions, offering ultra-sensitive sub-nanomolar detection.

Compare

An Electrochemical Detector (ECD) is a highly sensitive analytical instrument used in liquid chromatography systems, especially High-Performance Liquid Chromatography (HPLC) and Ion Chromatography (IC), for detecting electroactive species. It operates by measuring electrical current resulting from redox reactions of target analytes at the surface of an electrode. Designed for pharmaceutical, biomedical, environmental, and food safety laboratories, ECDs offer sub-nanomolar detection limits, making them ideal for analyzing neurotransmitters, vitamins, pesticides, phenols, sugars, and biomarkers.

Technical Specification of Electrochemical Detector

Mode DC ,PULSE and SCAN
Working Temperature 10-40℃(indoor use only)
Safety and EMC According to Ec-directives; Emission Group I Class A;cMETus approved
Sensors Up to 3 flow cells
Max Current Compensation(Autozero) – 25 nA-25 mA in DC and PULSE mode dependent on range setting
Oven +7°C above ambient to 60°C, accuracy 0.5°C, stability 0.1°C; accommodates column and flow cell(s)
Offset +50% to – 50% of max. output voltage, 5% steps
Analog Output (DAC) -1 to +1 V full scale (via 16-bit D/A converter)
Analog Output (I/E) -2.5 to +2.5 V full scale (unprocessed I/E con-verter signal)
DC mode
Range 10 pA-200 uA in 1. 2. 5 increments
Filter (ADF) 10-0.001 Hz in 1.2. 5 increments

RAW and OFF: for unprocessed data

Potential ( EC) -2.50V to +2.50V with 10 mV increments
Data Rate 1- 100 Hz in 1, 2, 5 increments, dependent on filter setting
Noise <2pA with dummy cell(load of 300 MΩ /470 pF) in 1 nA range, filter off, Ec+800mV and temperature of 35’C
PULSE Mode
Range 10 nA – 200 μA in 1, 2, 5 increments

OFF: for unprocessed data

Filter (ADF) 0.5 – 0.001 Hz in 1, 2, 5 increments
Potential (Ec) -2.50 V to + 2.50 V with 10 mV increments
Data Rate 1/(pulse duration) Hz
SCAN Mode
Range 10 nA – 200 μA in 1, 2, 5 increments
Potential (Ec) -2.50 V to + 2.50V with 10 mV increments
Data Rate 1 Hz Scan rate 1 – 100 mV/s in 1, 2, 5 increments
Others
Cycle Half, Full, Continuous
Dimensions 43(D)*22(W)*44(H)cm
Power requirements 100-240 VAC, 50/60 Hz, 260 VA, auto-sensing
Weight Max 14. 4 kg without flow cell and column
Category: Brand: