AST550 Fully-automatic Cryostat Microtome


Elevate your frozen tissue sectioning with our Fully-automatic Cryostat Microtome. Its advanced automation and precision make it the ideal choice for quick and consistent sample preparation, ensuring accurate diagnostics and research success.

Compare

Description of AST550 Fully-automatic Microtome Cryostat

  • It is designed under engineering and ergonomic concepts. All the components are processed by computer numerical control.
  • Retraction function prevents specimen damage from blade.
  • Total thickness calculation.
  • Low temperature refrigeration system with environment friendly refrigerant R404.
  • Adopt UV and O3 sterilize for 30 minutes every time.
  • Peltier specimen clamp refrigeration may be turned on or off manually.
  • Two defrosting methods: manual and automatic.
  • Two running ways: manual and automatic.
  • Emergency stop function: Press the red button to stop running when any emergency happens.
  • The high precision unit is enclosed outside the cryostat chamber to prevent it from thermal expansion and contraction, and keep minimum maintenance requirement.
  • Artificial intelligent interface is easy for user to learn and operate.
  • LCD screen can display all information including thickness, thickness sum, controlling time, temperature, timed on/off and so on.
  • Sleep Function: after activating this function, the temperature in cryostat chamber can be kept between -1℃ and – 10℃; After stopping this function, the temperature in freeze chamber will rise to working temperature within 30 minutes.
  • Control panel with locking function could avoid faulty operation.
  • Handwheel locking function.
  • Wide freezing shelf load 8 specimen cassettes simultaneously, and two of them are equipped with peltier to meet different temperatures.

Specification of AST550 Fully-automatic Microtome Cryostat

Freeze chamber minimum temperature -35℃±2℃
Freeze shelf minimum temperature -45℃±5℃
Lowest temperature of Peltier unit on freeze shelf -55℃±5℃
Lowest temperature of Peltier unit on Specimen Clamp -50℃±5℃(working time of Peltier is 15 minutes.)
Maximum specimen size 35×35mm
Specimen vertical stroke 48±1mm
Specimen horizontal stroke 15±0.2mm
Coarse feeding speed 300μm/s (slow) or 900μm/s(rapid)
Section thickness range 0μm~100μm adjustable;
0μm~3μm, in 0.5μm increments;
3μm~10μm, in 1μm increments;
10μm~20μm, in 2μm increments;
20μm~100μm, in 5μm increments;
Trimming thickness range 10μm~600μm adjustable;
10μm~50μm, in 5μm increments;
50μm~100μm, in 10μm increments;
100μm~600μm, in 50μm increments;
Retraction range 0~80μm adjustable, in 5μm increments
Voltage and frequency 220 V ± 10%, 50 HZ / 110 V ± 10%, 60 HZ
Nominal Power 1000 W
Weight about 130 KG
Dimensions(W×D×H) 805×715×1195mm