PmX-5050 is a fully stand-alone device for continuous analysis of particulate matter (PM) elemental composition, intended for online elemental analysis of heavy metals in the ambient air. Particulate matter’s presence in the air and its overall accumulation in the environment are the subjects of growing public concern, especially in the manufacturing facilities, plants, and factories surrounding areas. It is vital to measure and assess the PM concentration in the air to prevent any possible harm to natural resources, people residing in such areas and overall environmental deterioration of the region.
Benefits of PmX-5050
- Continuous analysis of PM 2.5 or PM 10 elemental concentration
- Full compliance with the US EPA IO-3.3 standard
- Up to 3 months of stand-alone continuous operation
- Automatic correction of ambient temperature, humidity and pressure influence
- Heated sampling stage for condensing prevention
- Simple calibration procedure and tape replacement
- Intuitive software interface
Technical Specifications
General
- Applicable standard: USA EPA IO-3.3
- Flow rate: (0~20) L/min
- Filter tape: PTFE
- Tape replacement interval: up to 3 months
- Power supply: AC 100V~240V, 50/60 Hz
- Power consumption: 200 W
- Dimensions (w/o sampling inlet): 500mm * 500mm * 450mm
- Weight (w/o sampling inlet): 40 kg
- External connection: USB, Ethernet
XRF unit
- X-Ray tube: 50W, 50kV, 3mA
- Primary beam filter changer: 5 positions, automatic
- X-Ray detector: SDD
- Sampling and analysis times: 3 ~ 1440 min
- Measurement range: 0 ~ 200 µ g/m3
- Detection limit: 0.1 ng/m3 for most elements
- Calibration samples: SS316 for energy calibration, NIST 2783 for concentration calibration, other samples(optional)
Environmental Conditions
- Ambient temperature: 0 ° C ~ 40 ° C
- Humidity: 0 ~ 80%
- Altitude: up to 1000m
There are no reviews yet.