The Hypoxia Workstation DY-TM is a high-capacity, precision-controlled hypoxia chamber engineered for advanced cell and tissue culture under low-oxygen environments. Designed for large-scale research, it features a 363-liter working volume and a 39-liter interlock, easily accommodating extensive experimental workflows. This oxygen-controlled incubator delivers highly accurate O₂ (0.1%–25%), CO₂ (0.1%–20%), and temperature and humidity control, making it ideal for studies in cancer biology, stem cell research, neuroscience, and cardiovascular pathology.
The DY-TM supports hypoxic, capnophilic, and optional anaerobic or hyperoxic modes, allowing flexible simulation of complex physiological conditions. Its tri-gas regulation system, real-time pressure monitoring, and programmable hypoxic cycles ensure a stable and repeatable research environment. A seamless entry system (SES) enables quick glove access without disturbing internal conditions, while quick sensor calibration and data logging streamline lab workflows.
Engineered for performance and reliability, the DY-TM also includes an internal socket and light, activated carbon sachet, and dual pressure relief system. Whether used for modeling tumor hypoxia or studying ischemic damage, the DY-TM hypoxia workstation offers the control and scalability needed for today’s most demanding biomedical research applications.
Features
- O2 Control: 0.1% – 23.0%, in 0.1% increments
- CO2 Control: 0.1% – 20.0%, in 0.1% increments
- One-touch calibration for O2 and CO2 sensor
- Temperature Control: 3°C above ambient up to 45°C, in 0.1°C increments
- Humidity Control: Room humidity to 85%
Optional Features
- Programmable Hypoxic Cycle: Automated adjustments of O2 and CO2 concentrations for efficient sample stimulation
- O2 Meter with Sensor: Continuously monitors internal O2 concentration
- Cable Port: Facilitates equipment integration within the workstation
- Remote Control with Alarm: Remote control and alarms for O2, CO2, temperature, humidity, gas cylinder pressure, and gas leaks
- Single Plate Entry System: Simplifies sample transfer
- Sidewall Pipette Holder: Convenient pipette holder for easy access
- Vac Line Connection Port: Simplifies waste disposal and cleaning processes
- Gas Sample Port: Convenient solution for gas extraction or detection
- High Temperature Chamber: Provides controlled high-temperature environments (up to 55°C) for experiments
- Micro-Pressure Chamber: Offers precise regulation of static pressure ranging from 0 to 35 KPa (0 to 5 PSI), with an adjustment accuracy of 0.1 KPa
- Anaerobic System: Combines hypoxia and anaerobic conditions in a single workstation
Specifications
Specification | Details |
---|---|
Internal Dimension (W/D/H mm) | 1100 / 570 / 550 |
External Dimension (W/D/H mm) | 1460 / 660 / 770 |
Gas Supplies | N₂ / Air |
Real-Time O₂ Control | Available |
O₂ Sensor Calibration | Available |
Direct Entry System | Available |
Interlock Purge | Available |
Internal Socket & Light | Available |
Real-time Data Graphing | Available |
USB Data Logging System | Available |
Three-Tiered User Password | Available |
Cylinder Low Pressure Alarm | Available |
Activated Carbon Sachet | Available |